Type II
Pure water

Your laboratory tasks determine which Omnia system is required. Pure water is needed not only for laboratory autoclaves and washing machines, but also for feeding to ultra pure water systems. You have the choice between only Type II water or additionally with Type I ultra pure water:

OmniaTap and OmniaLabED+ for both water qualities, Type I + II, as small or larger allrounder.

OmniaLabED, OmniaLabUP and OmniaLabDS are exclusively for pure water type Type II and optimized for various quantity requirements, purchasing and running costs, certification and safety requirements.



OmniaTap
Reinstwassersystem OmniaTap Der Allrounder.
For H2O pure types I + II.

 

OmniaTap is the ideal system when both pure water and ultrapure water are required, but in relatively small amounts. The ability to provide both types from a single system results from the combination of ultramodern purification technologies. These also make it possible to connect the system directly to a tap water. A press on the dispenser button activates dispensing of ultrapure water type I via the digital dispenser control. The recirculation of the pure water held in the installed 10 litre tank keeps it permanently at type II quality. The pure water tank has a second outlet for feeding downstream end users.


Features

Bulletpoint OptiFill dispenser is standard
Bulletpoint Pretreatment set for direct connection to tap water
Bulletpoint 10-litres pure water tank
Bulletpoint Tank volume display in percent
Bulletpoint Simple and economical filter replacement
Bulletpoint Leakage sensor is standard
Bulletpoint Ready-to-use, including filter cartridges


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on a
work surface



Flow chart

Flow chart OmniaTap


Specifications
OmniaTap
OmniaTap UV
OmniaTap UV/UF
Type II
Pure water performance l/h
at 15 °C
6 or 12
6 or 12
6 or 12
Conductivity µS/cm
0.067 – 0.1
0.067 – 0.1
0.067 – 0.1
Resistance MΩ x cm
15 – 10
15 – 10
15 – 10
Pure water tank
pressurized outlet
optional
optional
optional

Type I
Conductivit µS/cm
0.055
0.055
0.055
Resistance MΩ x cm
18.2
18.2
18.2
TOC-value* ppb
5 – 10
1 – 5
1 – 5
Dispensing performance l/min.
up to 2
up to 2
up to 1.6
Endotoxins* EU/ml
-
-
< 0.001
Particles**/ml
< 1
< 1
< 1
Bacteria** CFU/ml
< 0.1
< 0.1
< 0.1

Feedwater requirements
Tap water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
90-240/50-60
Connected load kW
0.1
0.1
0.1
Connector size mm
3/4"
3/4"
3/4"
Ambient temperature °C
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
Dimensions*** W x H x D mm
390 x 720 x 615
390 x 720 x 615
390 x 720 x 615
Weight kg
20
21
21

* in dependence on the feedwater quality       ** with sterilizing filter 0.2 µm       *** with OptiFill Dispenser


Article no.
System type*
Typical applications
18200051
OmniaTap 6
AAS, IC, ICP, buffers and media preparation
18200101
OmniaTap 12
AAS, IC, ICP, buffers and media preparation
18200052
OmniaTap 6 UV
Ultra-trace analysis, ICP-MS, HPLC, TOC
18200102
OmniaTap 12 UV
Ultra-trace analysis, ICP-MS, HPLC, TOC
18200053
OmniaTap 6 UV/UF
Life science and microbiology, cell culture media
18200103
OmniaTap 12 UV/UF
Life science and microbiology, cell culture media

* RO cartridge, ultrapure water cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter included


Accessories
19200300
Wall mount Omnia
19200056
Disinfection kit Omnia
19200057
Disinfectant Omnia – 3 pcs./pack
19200021
Pre-treatment unit OmniaTap –10“




 

OmniaLabED+
Pure and ultrapure water System OmniaLabED The big one.
For H2O pure types I + II.

 

OmniaLabED+ is the system of choice when both pure water and ultrapure water are needed for the entire laboratory. The system complies with international water standards such as ASTM, ISO 3696 and CLSI. The economy of it is maximized by the inclusion of a continuously self-regenerating electrodeionizer, without having to give any demanding analytical applications a pass. Further to this, each OmniaLabED+-system holds 100 litres of pure water type II ready for withdrawal in a storage tank that is equipped with quality recirculation. OmniaLabED+ is exactly right as pure water supplier to autoclaves and laboratory washing machines.


Features

Bulletpoint OptiFill dispenser is standard
Bulletpoint Continuous residual salts removal by electro-deionization
Bulletpoint 100 litre storage tank with recirculation and pressure outlet
Bulletpoint Tank volume display in percent
Bulletpoint Tank volume can be modularly increased
Bulletpoint Simple, cost-effective filter replacement
Bulletpoint Leakage sensor is standard


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on
a work surface

Tank fits spacesavingly under the bench-top

Tank fits space-
savingly under the
bench-top



Flow chart and ED-Module

CEDI-Modules


Specifications
OmniaLabED+20
OmniaLabED+40
OmniaLabED+70
Type II
Pure water performance l/h
at 15 °C
20
40
70
Conductivity µS/cm
0.067 – 1
0.067 – 1
0.067 – 1
Resistance* MΩ x cm
15 – 1
15 – 1
15 – 1
Silicate removal** %
99.9
99.9
99.9
Pure water tank
pressurized outlet
yes
yes
yes

Type I
Conductivity μS/cm
0.055
0.055
0.055
Resistance M Ω x cm
18.2
18.2
18.2
TOC-value* ppb (with UV-unit)
1 – 5
1 – 5
1 – 5
Dispensing performance dispenser l/min.
up to 2
up to 2
up to 2
Particles**/ml
< 1
< 1
< 1
Bacteria** CFU/ml
< 0.1
< 0.1
< 0.1

Feedwater requirements
Drinking water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
90-240/50-60
Connected load kW
0.25
0.25
0.25
Connector size mm
3/4"
3/4"
3/4"
Ambient temperature °C
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
Dimensions Tower***
W x H x D mm
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
Dimensions Base cabinet
tank mm
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
Weight kg
43
43
45

* in dependence on the feedwater quality       ** with sterilizing filter 0.2 µm       *** with OptiFill Dispenser


Article no.
System type*
Typical applications
18700020
OmniaLabED+20
Feedwater for autoclaves and laboratory washers
18700040
OmniaLabED+40
Feedwater for autoclaves and laboratory washers
18700070
OmniaLabED+70
Feedwater for autoclaves and laboratory washers

* RO cartridge, ultrapure water cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included


Accessories
16125000
Water softener MixMulti 32
19200020
Pre-treatment unit OmniaLab - 10“
19200050
UV Flow through disinfection – 254 nm
19200052
UV Unit for TOC reduction – 254 nm
16561201
External pressure booster pump SC 3000




 

OmniaLabED
Pure and ultrapure water System OmniaLabED The efficient one.
For H2O pure type II.

 

OmniaLabED is the efficient solution when high quality pure water Type II is required for the complete lab supply. It is compliant with international water standards, such as ASTM, ISO 3696, CLRW (CLSI), and the combination with continual self-regenerating electro-deionization brings maximized economy. Further to this, the OmniaLabED system holds 100 liters of pure water in a storage tank with quality recirculation, ready to supply lab equipment. OmniaLabED is the efficient one for supplying autoclaves, lab machines and ultra-pure water systems.


Features

Bulletpoint OptiFill dispenser is standard
Bulletpoint Continuous residual salts removal by electro-deionization
Bulletpoint 100 litre storage tank with recirculation and pressure outlet
Bulletpoint Tank volume display in percent
Bulletpoint Tank volume can be modularly increased
Bulletpoint Simple, cost-effective filter replacement
Bulletpoint Leakage sensor is standard


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on
a work surface

Tank fits spacesavingly under the bench-top

Tank fits space-
savingly under the
bench-top



Flow chart and ED-Module

CEDI-Modules


Specifications
OmniaLabED20
OmniaLabED40
OmniaLabED70
Type II
Pure water performance l/h
at 15 °C
20
40
70
Conductivity µS/cm
0.1 – 1
0.1 – 1
0.1 – 1
Resistance* MΩ x cm
10 – 1
10 – 1
10 – 1
TOC-value* ppb
< 30
< 30
< 30
Silicate removal* %
> 99
> 99
> 99
Dispensing performance dispenser l/min.
up to 2
up to 2
up to 2
Particles**/ml
< 1
< 1
< 1
Bacteria** CFU/ml
< 0.1
< 0.1
< 0.1
Pure water tank
pressurized outlet
yes
yes
yes

Feedwater requirements
Softened water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
90-240/50-60
Connected load kW
0.25
0.25
0.25
Connector size mm
3/4"
3/4"
3/4"
Ambient temperature °C
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
Dimensions Tower***
W x H x D mm
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
Dimensions Base cabinet
tank mm
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
Weight kg
41
41
43

* in dependence on the feedwater quality       ** with sterilizing filter 0.2 µm       *** with OptiFill Dispenser


Article no.
System type*
Typical applications
18700021
OmniaLabED20
Feedwater for autoclaves, laboratory washers
and ultrapure water systems
18700041
OmniaLabED40
Feedwater for autoclaves, laboratory washers
and ultrapure water systems
18700071
OmniaLabED70
Feedwater for autoclaves, laboratory washers
and ultrapure water systems

* RO cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included


Accessories
16125000
Water softener MixMulti 32
19200020
Pre-treatment unit OmniaLab - 10“
19200050
UV Tank disinfecting unit 254 nm
19200051
UV Flow through disinfection – 254 nm
16561201
External pressure booster pump SC 3000




 

OmniaLabUP
Pure Water System OmniaLabUP The constant one.
For H2O pure type II.

 

OmniaLabUP is the system of choice when you need a constant supply of high-quality water in laboratories. For this, OmniaLabUP holds 100 litres of type II pure water in reserve in a storage tank with quality recirculation. It is an optimal supplier to autoclaves, lab rinsing machines and ultrapure water systems. The water produced conforms to international medical technology water standards such as ASTM, ISO 3696 and CLRW (CLSI).


Features

Bulletpoint OptiFill Dispenser is standard
Bulletpoint 100 l tank with quality recirculation and pressure outlet
Bulletpoint Tank volume display in percent
Bulletpoint Tank volume can be modularly increased
Bulletpoint Simple, cost-effective filter replacement
Bulletpoint Leakage sensor is standard


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on a
work surface

Tank fits spacesavingly under the bench top

Tank fits space-
savingly under the
bench top



Flow chart

OmniaLab Up Flow chart class=


Specifications
OmniaLabUP20
OmniaLabUP40
Type II
Pure water performance l/h at 15 °C
20
40
Conductivity µS/cm
0.067 – 0.1
0.067 – 0.1
Resistance MΩ x cm
15 – 10
15 – 10
Dispensing performance dispenser l/min.
up to 2
up to 2
Pure water tank pressurized outlet
yes
yes
Particles*/ml
< 1
< 1
Bacteria* CFU/ml
< 0.1
< 0.1

Feedwater requirements
Softened or hardness-stabilized water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
Connected load kW
0.1
0.1
Connector size
3/4"
3/4"
Ambient temperature
+ 2 up to + 35 °C
+ 2 up to + 35 °C
Dimensions Tower* W x H x D mm
511 x 1520 x 575
511 x 1520 x 575
Dimensions Base cabinet tank mm
511 x 800 x 575
511 x 800 x 575
Weight kg
40
40

* with sterilizing filter 0.2 µm       ** with OptiFill Dispenser


Article no.
System type*
Typical applications
18600020
OmniaLabUP20
Feedwater for autoclaves and laboratory washers
18600040
OmniaLabUP40
Feedwater for autoclaves and laboratory washers

* RO cartridge, ultrapure water cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included


Accessories
19200020
Pre-treatment unit OmniaLab – 10“
19200050
UV tank disinfection unit 254 nm
19200051
UV Flow through disinfection 254 nm
16561201
External pressure booster pump SC 3000




 

OmniaLabDS
OmniaLabDS The reliable one.
For H2For H2O pure pure Type II + CLRW (CLSI)

 

When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.


Features

Bulletpoint OptiFill dispenser is standard
Bulletpoint 100 l tank with quality recirculation and pressure outlet
Bulletpoint Tank volume display in percent
Bulletpoint Tank volume can be modularly increased
Bulletpoint Simple, cost-effective filter replacement
Bulletpoint Leakage sensor is standard
Bulletpoint Emergency supply (optional)
Bulletpoint Degassing unit (optional)


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on
a work surface

Tank fits spacesavingly under the bench-top

Tank fits space-
savingly under the
bench-top



Flow chart

Flow chart OmniaLab DS


Specifications
OmniaLabDS20
OmniaLabDS40
OmniaLabDS60
OmniaLabDS80
Type II + CLRW (CLSI) DIN EN 285 + ISO EN 15883
Pure water performance l/h
at 15 °C
20
40
60
80
Conductivity μS/cm
0.1 – 1.0
0.1 – 1.0
0.1 – 1.0
0.1 – 1.0
Resistance M Ω x cm
10 – 1
10 – 1
10 – 1
10 – 1
Dispensing performance dispenser l/min.
up to 2
up to 2
up to 2
up to 2
Pure water tank pressurized outlet
yes
yes
yes
yes
Particles*/ml
< 1
< 1
< 1
< 1
Bacteria* CFU/ml
< 0.1
< 0.1
< 0.1
< 0.1

Feedwater requirements
Softened or hardness-stabilized water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
90-240/50-60
90-240/50-60
Connected load kW
0.1
0.1
0.1
0.1
Connector size
3/4"
3/4"
3/4"
3/4"
Ambient temperature °C
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
Dimensions Tower** W x H x D mm
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
Dimensions Base cabinet tank mm
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
Weight kg (without polishing cartridge)
39
39
40
40

* with sterilizing filter 0.2 μm       ** with OptiFill Dispenser


Article no.
System type*
Typical applications
18800020
OmniaLabDS20
Feedwater for analyzers, autoclaves and laboratory washers
18800040
OmniaLabDS40
Feedwater for analyzers, autoclaves and laboratory washers
18800060
OmniaLabDS60
Feedwater for analyzers, autoclaves and laboratory washers
18800080
OmniaLabDS80
Feedwater for analyzers, autoclaves and laboratory washers

* RO cartridge, stainless steel polishing cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included


Accessories
19200020
Pre-treatment unit OmniaLab – 10“
19200050
UV tank disinfection unit 254 nm
19200051
UV Flow through disinfection 254 nm
12280050
Replacement/second polishing cartridge type DS 2800 RV
19200040
Emergency supply
19200041
Degassing unit
16561201
External pressure booster pump SC 3000