OmniaLabDS
OmniaLabDS The reliable one.
For H2For H2O pure pure Type II + CLRW (CLSI)

 

When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.


Features

Bulletpoint OptiFill dispenser is standard
Bulletpoint 100 l tank with quality recirculation and pressure outlet
Bulletpoint Tank volume display in percent
Bulletpoint Tank volume can be modularly increased
Bulletpoint Simple, cost-effective filter replacement
Bulletpoint Leakage sensor is standard
Bulletpoint Emergency supply (optional)
Bulletpoint Degassing unit (optional)


One hand operation

One hand
operation

Easy water dispensing

Easy water
dispensing

Flexible on a work surface

Flexible on
a work surface

Tank fits spacesavingly under the bench-top

Tank fits space-
savingly under the
bench-top



Flow chart

Flow chart OmniaLab DS


Specifications
OmniaLabDS20
OmniaLabDS40
OmniaLabDS60
OmniaLabDS80
Type II + CLRW (CLSI) DIN EN 285 + ISO EN 15883
Pure water performance l/h
at 15 °C
20
40
60
80
Conductivity μS/cm
0.1 – 1.0
0.1 – 1.0
0.1 – 1.0
0.1 – 1.0
Resistance M Ω x cm
10 – 1
10 – 1
10 – 1
10 – 1
Dispensing performance dispenser l/min.
up to 2
up to 2
up to 2
up to 2
Pure water tank pressurized outlet
yes
yes
yes
yes
Particles*/ml
< 1
< 1
< 1
< 1
Bacteria* CFU/ml
< 0.1
< 0.1
< 0.1
< 0.1

Feedwater requirements
Softened or hardness-stabilized water according to DIN 2000
Feedwater temperature °C
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Manganese and iron content mg/l
< 0.05
< 0.05
< 0.05
< 0.05
Free chlorine content mg/l
< 0.1
< 0.1
< 0.1
< 0.1
Silt density index (SDI)
max. 3
max. 3
max. 3
max. 3

Technical data
Operating pressure bar
2 – 6
2 – 6
2 – 6
2 – 6
Supply voltage Volt/Hz
90-240/50-60
90-240/50-60
90-240/50-60
90-240/50-60
Connected load kW
0.1
0.1
0.1
0.1
Connector size
3/4"
3/4"
3/4"
3/4"
Ambient temperature °C
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
+ 2 up to + 35
Dimensions Tower** W x H x D mm
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
511 x 1520 x 575
Dimensions Base cabinet tank mm
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
511 x 800 x 575
Weight kg (without polishing cartridge)
39
39
40
40

* with sterilizing filter 0.2 μm       ** with OptiFill Dispenser


Article no.
System type*
Typical applications
18800020
OmniaLabDS20
Feedwater for analyzers, autoclaves and laboratory washers
18800040
OmniaLabDS40
Feedwater for analyzers, autoclaves and laboratory washers
18800060
OmniaLabDS60
Feedwater for analyzers, autoclaves and laboratory washers
18800080
OmniaLabDS80
Feedwater for analyzers, autoclaves and laboratory washers

* RO cartridge, stainless steel polishing cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included


Accessories
19200020
Pre-treatment unit OmniaLab – 10“
19200050
UV tank disinfection unit 254 nm
19200051
UV Flow through disinfection 254 nm
12280050
Replacement/second polishing cartridge type DS 2800 RV
19200040
Emergency supply
19200041
Degassing unit
16561201
External pressure booster pump SC 3000