Close

Type II + CLRW (CLSI) + DIN EN 285 + EN ISO 15883

Pure water for clinical laboratory

These water treatment systems for feeding analytical instruments generate pure water for clinical laboratory (Type CLRW) while using less water and furthermore reducing substantially tap water costs of laboratories. Water is the most common liquid reagent in a clinical analyzer. Therefore, selecting the right water treatment system is the key for a seamless and uninterruptible laboratory operation.

OmniaLabDS is the perfect solution when safety is first priority and the quality of the purification decides the quality of results.

Details

OmniaLabDS

The reliable one.
For H2O pure types II + CLRW (CLSI) + DIN EN 285 + EN ISO 15883.

When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.

Features

  • OptiFill dispenser is standard
  • 100 l tank with quality recirculation and pressure outlet
  • Tank volume display in percent
  • Tank volume can be modularly increased
  • Simple, cost-effective filter replacement
  • Leakage sensor is standard
  • Emergency supply (optional)
  • Degassing unit (optional)


One hand operation
One hand
operation
Easy water dispensing
Easy water
dispensing
Flexible on a work surface
Flexible on
a work surface
Tank fits space-savingly under the<br />
bench-top
Tank fits space-savingly under the bench-top

Specifications

OmniaLabDS20

OmniaLabDS40

OmniaLabDS60

OmniaLabDS80
Type II + CLRW (CLSI) DIN EN 285 + ISO EN 15883
Pure water performance l/h
at 15 °C
20406080
Conductivity μS/cm0.1 – 1.00.1 – 1.00.1 – 1.00.1 – 1.0
Resistance M Ω x cm10 – 110 – 110 – 110 – 1
Dispensing performance dispenser l/min.up to 2up to 2up to 2up to 2
Pure water tank pressurized outletoptionaloptionaloptionaloptional
Particles*/ml< 1< 1< 1< 1
Bacteria* CFU/ml< 0.1< 0.1< 0.1< 0.1

Feedwater requirements
Softened or hardness-stabilized water according to DIN 2000
Feedwater temperature °C+ 2 up to 35+ 2 up to 35+ 2 up to 35+ 2 up to 35
Manganese and iron content mg/l< 0.05< 0.05< 0.05< 0.05
Free chlorine content mg/l< 0.1< 0.1< 0.1< 0.1
Silt density index (SDI)max. 3max. 3max. 3max. 3

Technical data
Operating pressure bar2 – 62 – 62 – 62 – 6
Supply voltage Volt/Hz90-240/50-6090-240/50-6090-240/50-6090-240/50-60
Connected load kW0.10.10.10.1
Connector sizeR 3/4"R 3/4"R 3/4"R 3/4"
Ambient temperature °C+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
+ 2 up to 35
Dimensions

Tower** W x H x D mm
511 x 1520 x 575511 x 1520 x 575511 x 1520 x 575511 x 1520 x 575
Dimensions

Base cabinet tank mm
511 x 800 x 575511 x 800 x 575511 x 800 x 575511 x 800 x 575
Weight kg (without polishing cartridge)
39394040
* with sterilizing filter 0.2 μm
** with OptiFill Dispenser

Article no.

System type*

Typical applications
18800020OmniaLabDS20Feedwater for analyzers, autoclaves and laboratory washers
18800040OmniaLabDS40Feedwater for analyzers, autoclaves and laboratory washers
18800060OmniaLabDS60Feedwater for analyzers, autoclaves and laboratory washers
18800080OmniaLabDS80Feedwater for analyzers, autoclaves and laboratory washers

* RO cartridge, stainless steel polishing cartridge, sterile filter capsule 0.2 μm, sterile overflow and sterile vent filter + CO2 absorber included

Accessories
19200020Pre-treatment unit OmniaLab – 10“
19200050UV tank disinfection unit 254 nm
19200051UV Flow through disinfection 254 nm
12280050Replacement/second polishing cartridge type DS 2800 RV
19200040Emergency supply
19200041Degassing unit
16561201External pressure booster pump SC 3000

Fließschema OmniaLabDS

Media

Omnia
3D-Video

OMNIA Produktvideo

Brochure pure and ultrapure water systems